ARUP's Laboratory Test Directory

Troponin I : 0090613

Mnemonic: TNI

Methodology: Chemiluminescent Immunoassay
Performed: Sun-Sat
Reported: Within 24 hours
Specimen Required: Collect:  One 3 mL plasma separator tube.

Transport:  1 mL plasma, frozen.  (Min: 0.5 mL)

Pediatric Collection/Transport:  0.5 mL plasma, frozen.

Remarks:  Avoid serum samples.  Remove plasma from cells ASAP.  Samples must be free of particulate matter including fibrin which can interfere with the assay.  If testing cannot be performed within 48 hours of collection, ship frozen.  For plasma samples: Avoid transferring material from the white blood cell/platelet layer located just above the red blood cells. 

Unacceptable Conditions:  Samples exposed to multiple freeze/thaw cycles.

Stability:  Ambient: 8 hours; Refrigerated: 2 days; Frozen: 6 months
Reference Interval:
0.0-0.05 ng/mL
Interpretive Data: 0.05 ng/mL or less : Negative - repeat testing in four to six hours if clinically indicated.
0.06-0.49 ng/mL: Suspicious for myocardial injury. Serial measurements may be necessary to confirm or exclude the diagnosis of acute coronary syndrome.  Repeat testing in four to six hours if indicated.
0.50 ng/mL or greater : Consistent with myocardial injury. Clinical and laboratory correlation recommended.
CPT Code(s): 84484
 
 

 

 

 
All ARUP Sites:        www.aruplab.com     ·     www.arupconsult.com     ·     www.arup.utah.edu     ·     www.childx.org     ·     www.utahblood.org