ARUP's Laboratory Test Directory

Hepatitis C Virus Antibody by CIA : 2002483

Mnemonic: HCV AB

Methodology: Chemiluminescent Immunoassay
Performed: Sun-Sat
Reported: Within 24 hours
Specimen Required: Collect: One 4 mL serum separator tube.

Storage/Transport Temperature: 1 mL serum at 2-8°C. (Min: 0.5 mL) Submit specimen in an ARUP Standard Transport Tube.

Remarks: Potassium EDTA plasma is also acceptable. Separate serum from cells ASAP.

Unacceptable Conditions: Specimens containing particulate material. Severely hemolyzed or lipemic specimens. Heat inactivated specimens. Heparinized plasma.

Stability (collection to initiation of testing): After separation from cells: Ambient: Unacceptable; Refrigerated: 1 week; Frozen: Indefinitely (avoid freeze/thaw cycles)
Reference Interval:
Negative
Interpretive Data: Negative: 0.79 IV or less
Equivocal: 0.80 to 0.99 IV
Low Positive: 1.00 to 10.99 IV
High Positive: 11.00 IV or greater
IV = S/C

ARUP intends use of this assay for clinical diagnosis. This assay should not be used for blood donor screening, associated re-entry protocols, or for screening Human Cell, Tissues and Cellular and Tissue-Based Products (HCT/P).

Note: The supplemental anti-HCV Recombinant Immunoblot Assay (RIBA) is recommended for specimens with low positive anti-HCV screening results. For high positive anti-HCV screening results, ARUP recommends collecting a new specimen and testing for Hepatitis C RNA by PCR or bDNA.
CPT Code(s): 86803
 
 

 

 

 