ARUP's Laboratory Test Directory

QuantiFERON®-TB Gold In Tube & Mycobacterium tuberculosis Antibody, IgG : 2001627

Mnemonic: TB PAN

Methodology: Cell Culture/Quantitative Enzyme-Linked Immunosorbent Assay
Performed: Thu
Reported: 1-8 days
Specimen Required: Collect: At altitudes from sea level to 2,650 feet: Whole blood directly into each of the three tubes in the QuantiFERON®-TB (Standard) Gold In-Tube collection kit (ARUP supply #45112). Available online through eSupply using ARUP Connect™ or contact ARUP Client Services at (800) 522-2787.
At altitudes from 2,650-3,350 feet:
Collect 3 mL whole blood using a syringe. Transfer 1 mL (Min: 0.8 mL, Max 1.2 mL) to each of the three tubes in QuantiFERON®-TB (Standard) Gold In-Tube collection kit. (ARUP supply #45112).
At altitudes from 3,350-6,150 feet:
Whole blood directly into each of the three tubes in the QuantiFERON®-TB (High Altitude)Gold In-Tube collection kit (ARUP supply #46261). Available online through eSupply using ARUP Connect™ or contact ARUP Client Services at (800) 522-2787.
At altitudes above 6,150 feet:
Collect 3 mL whole blood using a syringe. Transfer 1 mL (Min: 0.8 mL, Max 1.2 mL) to each of the three tubes in QuantiFERON®-TB (High Altitude) Gold In-Tube collection kit.
Tubes should be between 17°C and 25°C (63°F to 77°F) at the time of blood filling.


Specimen Preparation: Immediately after filling tubes, shake them ten (10) times just firmly enough to ensure the entire inner surface of the tube is coated with blood, to solubilize antigens on tube walls. Over-energetic shaking may cause gel disruption and could lead to aberrant results.
After mixing,
incubate upright at 37 ° C for 16-24 hours (within 16 hours of collection).
After incubation,
centrifuge for 15 minutes at 2000-3000 RCF (g).
Transport plasma in each of the 3 original collection tubes.


Storage/Transport Temperature: Refrigerated.

Remarks: Further information on mixing QuantiFERON-TB Gold In-Tube collection tubes and an instructional video can be found at: http://www.aruplab.com/Specimen-Handling/SpecialSpecimenCollection/quantiferon.jsp

Unacceptable Conditions: Room temperature specimens. Specimens that have not been centrifuged.

Stability (collection to initiation of testing): After incubation and centrifugation: Ambient: 2 hours; Refrigerated: 4 weeks; Frozen: 4 months

Reference Interval:
 
Test Number Components Reference Interval
0051729QuantiFERON®-TB Gold In Tube  
 
QuantiFERON®-TB Gold In Tube
Effective November 19, 2012
Components Reference Interval
Quantiferon TB Gold In-Tube Negative
Quantiferon TB minus NIL 0.34 IU/mL or less
Quantiferon MITOGEN minus NIL No reference interval
Quantiferon NIL No reference interval
0051698Mycobacterium tuberculosis Antibody, IgG by ELISA 0.424 OD or less: Negative - No significant level of detectable IgG antibody to Mycobacterium tuberculosis.

0.425-0.500 OD: Equivocal  - Repeat testing in 10-14 days may be helpful.

0.501 OD or greater: Positive  - IgG antibody to Mycobacterium tuberculosis detected, which may indicate acute or active infection.

Interpretive Data: Refer to individual components for further interpretive data.
Note: Refer to individual components for further notes.
CPT Code(s): 86480; 86609